Process Measurement Company & MadgeTech Bring Oven Mapping Solution to Medical Device Manufacturer

OVERVIEW

Process Measurement Company (PMC) was recently approached by a medical device manufacturer to conduct an oven mapping study. A report of these results is requested on an annual basis to prove compliance with government and company policy. The customer had certain temperature, tolerance, and timing requirements. To accomplish the oven mapping study, PMC used MadgeTech’s Titan S8 portable data acquisition logger.

THE CHALLENGE

The company needed their clean process benchtop oven to be validated at a temperature of 150 °C with a tolerance of +/- 1.5 °C with readings required every 30 seconds over an hour. This type of oven is typically used for packaging processes, sterilization, and drying.

THE SOLUTION

MadgeTech Titan S8To efficiently perform the oven mapping, PMC used MadgeTech’s Titan S8 portable data acquisition logger. Equipped with eight channel inputs, out technician placed a Type K thermocouple at the top and bottom of each corner of the oven. With the Titan S8, the technician was able to map the eight points while viewing the results in real-time, graph, and tabular view.

THE RESULT

Over the course of the study, 976 readings were collected, downloaded onto a USB and uploaded into a Microsoft Excel file for analysis. Thanks to MadgeTech’s Titan S8, results showed that all eight points of the oven were operating properly at or slightly above 150 °C and Process Measurement was able to issue a certificate of conformance.

Read the full case study from MadgeTech on their website.

ABOUT PMC

Process Measurement Company is a an ISO/ IEC accredited calibration lab, offering equipment and instrument calibration services including electrical, dimensional, physical, and mechanical for a wide variety of industries. In addition to drop-off services, PMC also offers onsite calibration services. To request a quote, contact us today.

2018-08-30T15:57:46-05:00August 30th, 2018|